7 rls dos, 1 rls dos - overview, Rls dos – Metrohm tiamo 2.4 Manual User Manual
Page 1228
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1212
■■■■■■■■
tiamo 2.4
Solution
Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.
Entry
24 characters
Selection
'Solution name' | not defined
Default value
not defined
not defined
No tests will be carried out.
5.6.6.7
RLS DOS
5.6.6.7.1
RLS DOS - Overview
Dialog window: Method
▶ RLS DOS ▶ Properties... ▶ RLS DOS - 'Command name'
Command to release a dosing device for using it in other parallel run-
ning methods. It removes the reservation made by the current method.
Devices
This command can be executed with the following devices:
Sample Processor: 730, 774, 778, 789, 814, 815, 864, 874
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command RLS DOS are configured in the follow-
ing dialog window:
■
Command variables
The following command variables are generated in the method run by the
command RLS DOS and can be used in formulas under the designation
'Command name.Variable designation':