5 Method
tiamo 2.4
■■■■■■■■
863
5.6.3.8.4
STAT - Track call
Dialog window: Method
▶ STAT/STAT pH/STAT U ▶ Properties... ▶ Monitor-
ing
▶ [New]/[Properties] ▶ Track call #
Monitoring
Selection of the quantity to be monitored. When its limit values are viola-
ted a track is started.
Selection
Measured value | Dosing rate | Temperature |
Any
Default value
Measured value
Any
Means any of the three quantities.
Limit exceeding
Selection of the exceeded limit for which a track is to be started.
Selection
Lower limit | Upper limit | Any | OK
Default value
Any
Any
The track will be started if either the upper or the lower limit is viola-
ted.
OK
The track will be started when the monitored quantities are once again
within the limit values (including hysteresis).
Track name
Selection of the track that is to be started automatically.
Selection
'Track name'
NOTE
If a track is called for that is already running then it will be allowed to
finish and then restarted when it is free again.
5.6.3.8.5
Tandem dosing
Menu item: Method
▶ STAT/ADD/DOS ▶ General/Hardware
If the Tandem dosing check box is enabled for the commands STAT,
ADD or DOS, then a second dosing device can be defined in order to
enable uninterrupted dosing. Dosing is carried out with a combination of
two dosing devices so that the second dosing device is dosing while the
first one is being filled and vice versa.