Metrohm tiamo 2.4 Manual User Manual
Page 984
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
968
■■■■■■■■
tiamo 2.4
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead
If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
Device types | Titrando
Default value
Titrando
Sensor
Measuring input
Selection of the measuring input the sensor is connected to.
Titrando (without 888, 890)
Selection
1 | 2
Default value
1
851, 855, 867, 888, 890
Selection
1
Default value
1
852
Selection
2
Default value
2
Sensor
Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.
Selection
Sensor name | Metal electrode
Default value
Metal electrode
I(pol)
The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.
Titrando, 855, 867
Input range
-125.0 - 125.0 µA (Increment: 0.5)
Default value
5.0 µA