5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1238
■■■■■■■■
tiamo 2.4
5.6.7.9.2
FLOW - Properties
Dialog window: Method
▶ FLOW ▶ Properties... ▶ FLOW - 'Command name'
Dialog window: FLOW - 'Command name'
Command name
Name of the command.
Entry
25 characters
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Shows the device type. If an instrument is selected in the Device name
field, the Device type field can no longer be edited and the instrument
type belonging to the instrument is displayed.
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
774 Oven Sample Processor | 874 USB Oven
Sample Processor
Default value
874 USB Oven Sample Processor
Gas
Gas type
Gas type of the carrier gas.
Selection
Air | Nitrogen | Other gas
Default value
Air
Gas flow factor
Factor for the correct determination of the gas flow. This parameter is dis-
played only if Gas type = Other gas. For the 774 Oven Sample Pro-
cessor, the gas flow factor values for commonly used carrier gases are lis-