5 Method
tiamo 2.4
■■■■■■■■
513
■
Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.
■
Additional measured values
Definition of additional measured values of other measuring com-
mands, which are saved as additional columns in the measuring point
list.
Command variables
The following command variables are generated by the DET Ipol com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identification
Description
.BP{x}.ERC
ERC or first derivative for the break point × (1 - 9)
.BP{x}.MEA
Measured value for the break point × (1 - 9) in the
unit of the measured value
.BP{x}.TEM
Temperature for the break point × (1 - 9) in °C
.BP{x}.TIM
Time for the break point × (1 - 9) in s
.BP{x}.VOL
Volume for the break point × (1 - 9) in mL
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.CONC
Concentration of the solution used for the command
.CYL
Cylinder volume of the exchange or dosing unit used
for the command
.DBL
Total duration for the processing of the command in
s
.DSC
Time for processing all start conditions in s
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.EP{x}.ERC
ERC for the endpoint × (1 - 9)
.EP{x}.MEA
Measured value for the endpoint × (1 - 9) in the unit
of the measured value
.EP{x}.MEP
Number of endpoints in the window × (1 - 9); 1 = 1
endpoint, 2 = 2 or more endpoints, 3 = EP cor-
rected with Autodrift, 4 = EP corrected with
manual drift