5 Method
tiamo 2.4
■■■■■■■■
1175
Tandem dosing
on | off (Default value: off)
If this option is enabled, then dosing will proceed continuously with two
dosing devices in combination. The second dosing device is dosing while
the first one is being filled and vice-versa. This parameter is displayed only
for Titrandos and 855.
Dosing device
Selection of the second dosing device (exchange or dosing unit) to be
used for the dosing if the first dosing device is not available.
Titrando
Selection
1 | 2 | 3 | 4
Default value
2
855
Selection
1 | 2 | 3
Default value
2
Solution
Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked.
Titrando, 855
Selection
Solution name | not defined
Default value
not defined
not defined
No tests will be carried out.
Filling rate
Speed with which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the
exchange unit or dosing device used. If the entered filling rate is too high
for the selected dosing device, it will automatically be reduced during dos-
ing to the largest possible value.
Titrando, 855
Input range
0.01 - 166.00 mL/min
Default value
maximum mL/min