5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
512
■■■■■■■■
tiamo 2.4
Time
Input range
0.0 - 999999.9 s
Volume
Input range
0.00000 - 9999.99 mL
5.6.3.2.4
DET Ipol
5.6.3.2.4.1
DET Ipol - Overview
Dialog window: Method
▶ DET Ipol ▶ Properties... ▶ DET Ipol - 'Command name'
Command for Dynamic equivalence point titrations with voltametric
measurement (selectable polarization current).
Devices
This command can be executed with the following instruments:
Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907
Titrino: 716, 721, 736, 751, 785, 794, 798, 799
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the DET Ipol command are set on the following
seven tabs:
■
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
■
Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.
■
Titration parameters
Parameters for the run of the titration.
■
Stop conditions
Definition of conditions which cause the titration to stop.
■
Potentiometric evaluation
Parameters for the potentiometric evaluation of titration curves.