4 brc - control parameters, Control parameters – Metrohm tiamo 2.4 Manual User Manual
Page 867
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
851
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.
Input range
0 - 999999 s
Default value
0 s
Max. waiting time
If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed.
Input range
0 - 999999 s
Default value
15 s
Pause
Pause
Waiting time before the start of the titration.
Input range
0 - 999999 s
Default value
0 s
5.6.3.7.4
BRC - Control parameters
Tab: Method
▶ BRC ▶ Properties... ▶ Control parameters
Command name
Name of the command.
Entry
25 characters
Endpoint
EP with
Measured value for the endpoint.
Input range
-1250.0 - 1250.0 mV
Default value
200.0 mV
Titration rate
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.
Selection
slow | optimal | fast | User
Default value
optimal