2 cal loop ph - properties – Metrohm tiamo 2.4 Manual User Manual
Page 1156
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1140
■■■■■■■■
tiamo 2.4
Command variables
The following command variables are generated by the CAL LOOP pH
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':
Identifi-
cation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
never been started
.ENP
Electrode zero point calculated from the calibration
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command has
never ended
.LCO
Loop counter = current number of completed runs, both
of Repeat loops and of While loops
.LST
Start time of the loop command (Date/Time)
.SLO
Electrode slope calculated from the calibration
.VAR
Variance of the calculated result of the calibration in the
selected unit
5.6.5.12.2
CAL LOOP pH - Properties
Dialog window: Method
▶ CAL LOOP pH ▶ Properties... ▶ CAL LOOP pH - 'Com-
mand name'
The number of buffers and the buffer type are defined in the CAL LOOP
pH - 'Command name' dialog window.
Command name
Name of the command.
Entry
25 characters
Buffers
Number of buffers
Selection of the number of buffers with which the pH calibration is to be
carried out.
Selection
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9
Default value
2
Buffer type
Selection of a predefined buffer series for the pH calibration.