7.9 774 Oven Sample Processor
■■■■■■■■■■■■■■■■■■■■■■
1534
■■■■■■■■
tiamo 2.4
Max. temperature
With the 774 Oven Sample Processor display of the maximum tempera-
ture set on the device only.
With the 774 Oven Sample Processor display only.
Input range
50 - 250 °C
Default value
250 °C
Temperature correction
Enter the temperature correction. This parameter directly affects the tem-
perature control and allows it to be adjusted. In this way any temperature
difference that may occur between the oven and sample temperature can
be compensated.
With the 774 Oven Sample Processor display of the temperature correc-
tion set on the device only.
Input range
-10 - +10 °C
Default value
0 °C
7.9.7
Properties - Gas
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Gas
On the tab Gas the parameters for the gas flow on the 774 Oven Sam-
ple Processor are shown. The parameters cannot be edited, they are
only read off by the device.
Unit for gas flow
Shows the unit selected at the device for the gas flow display (mL/min or
L/h).
Minimum gas flow
Shows the lower warning limit set on the device for the gas flow.
Maximum gas flow
Shows the upper warning limit set on the device for the gas flow.
NOTE
If the minimum gas flow is undercut or the maximum gas flow is
exceeded then an error message will appear.