5 cal cond - data conductivity standard – Metrohm tiamo 2.4 Manual User Manual
Page 1124
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1108
■■■■■■■■
tiamo 2.4
Template
If this option is selected, then a conductivity standard can be selected,
and its data will be used to calculate the temperature coefficient for
the specified reference temperature.
Parameters for input
If the option Input has been selected, then the following parameters will
be displayed:
Reference temperature
Input of a temperature for which a calibration point of the conductivity
standard is known. It should be as close as possible to the temperature at
which measurements are carried out.
Input range
-20.0 - 150.0°C
Default value
20.0
Conductivity
Input of the conductivity of the standard at the reference temperature.
Input range
0.00000 - 2000.00 mS/cm
Default value
11.66 mS/cm
Temperature coefficient
Temperature coefficient of the standard with the reference temperature.
Input range
0.00 - 9.99 %/°C
Default value
2.07 %/°C
Parameters for template
If the option Input has been selected, then the following parameters will
be displayed:
Name standard
List box with conductivity standards. The data of one of these standards is
used to calculate the temperature coefficient for the specified reference
temperature.
Selection
'Conductivity standard'
5.6.5.4.5
CAL Cond - Data Conductivity standard
Dialog window: Method
▶ CAL Cond ▶ Data Conductivity standard
Current temperature
Display of the currently measured temperature of the conductivity stand-
ard.