3 kft upol - start conditions, Start conditions – Metrohm tiamo 2.4 Manual User Manual
Page 829
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
813
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando
Input range
-15 - 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos.
Switch on/off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command. This parameter is displayed only for 751, 785, 784,
795 and 799.
5.6.3.5.3.3
KFT Upol - Start conditions
Tab: Method
▶ KFT Upol ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Entry
25 characters
The start conditions are processed in the sequence shown before the titra-
tion is started.