5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
830
■■■■■■■■
tiamo 2.4
Designation
Description
.FP{x}.QTY
Measured value (water) for the fixed endpoint × (1…9)
in µg
.FP{x}.TEM
Temperature for the fixed endpoint x (1…9) in °C
.FP{x}.TIM
Time in s until the fixed endpoint × (1…9) is reached
.IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.CAx
Calculated value × (1...3) for the last measuring point
on the measuring point list
.LP.CHA
Charge for the last measuring point of the measuring
point list in mA s
.LP.DVT
dV/dt for the last measuring point on the measuring
point list (SET, STAT, DOS) or drift for the last measur-
ing point on the measuring point list in µg/min (KFC)
.LP.EXx
External value × (1...3) for the last measuring point of
the measuring point list
.LP.IGE
Current pulsed current for the last measuring point in
the measuring point list in mA
.LP.MEA
Measured value for the last measuring point of the
measuring point list in the unit of the measured value
.LP.QTY
Measured value (water) for the last measuring point in
the measuring point list in µg
.LP.TEM
Temperature for the last measuring point of the mea-
suring point list in °C
.LP.TIM
Time in s until the last measuring point in the measur-
ing point list is reached
.LP.UGE
Voltage at the generator electrode for last measuring
point of the measuring point list in V; in case of meas-
urements with 756, 831: 0 = not defined, 1 = <14
V, 2 = 14…28 V, 3 = >28 V
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off
.NMP
Number of measuring points in measuring point list