5 Method
tiamo 2.4
■■■■■■■■
717
5.6.3.4.3.10
SET pH - Fixed endpoint evaluation
Dialog window: Method
▶ SET pH ▶ Properties... ▶ Additional evaluations ▶ Fixed
endpoint evaluation
▶ [New/[Properties] ▶ Fixed endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Input range
-20.000 - 20.000 pH
Time
Input range
0.0 - 999999.9 s
Volume
Input range
0.00000 - 9999.99 mL
5.6.3.4.4
SET U
5.6.3.4.4.1
SET U - Overview
Dialog window: Method
▶ SET U ▶ Properties... ▶ SET U - 'Command name'
Command for endpoint titrations with potentiometric voltage measure-
ment.
Devices
This command can be executed with the following instruments:
Titrando: 808, 809, 835, 836, 841, 842, 857, 888, 901, 902, 904, 905,
906, 907
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798,
799
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
without conditioning
with conditioning