5 kft ipol - titration parameters, Titration parameters – Metrohm tiamo 2.4 Manual User Manual
Page 814

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
798
■■■■■■■■
tiamo 2.4
Stop time
The titration is canceled when the absolute stop time defined here has
elapsed since the start of the titration. This parameter is displayed only if
Stop criterion = Time and if Delay time = inf..
Titrino
Input range
1 - 999,999 s
5.6.3.5.2.5
KFT Ipol - Titration parameters
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the running of the titration are defined on this tab.
Titration direction
Selection of the titration direction.
Selection
+ | - | auto
Default value
auto
+
positive measured value change
-
negative measured value change
auto
The titration direction is determined automatically from the initial mea-
sured value and the set endpoint.
Extraction time
The titration is not canceled until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, for instance, for the titration of sparingly soluble sam-
ples.
Input range
0 - 999,999 s
Temperature
Manually entered titration temperature. If a temperature sensor is con-
nected and the Temperature measurement is set on the General/
Hardware tab under Sensor to automatic or continuous, then the
temperature will be measured continuously. This value is used for temper-
ature correction in pH measurements.