5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
516
■■■■■■■■
tiamo 2.4
Identification
Description
.SVA
Start volume absolute (volume that was added
according to the start condition "start volume") in mL
.SVM
Start volume measured value (volume that was added
according to the start condition "start measured
value") in mL
.SVS
Start volume slope (volume that was added according
to the start condition "start slope") in mL
.SVT
Total start volume (volume that was added according
to all three start conditions) in mL
.TITER
Titer value of the solution used for the command
5.6.3.2.4.2
DET Ipol - General/Hardware
Tab: Method
▶ DET Ipol ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the instrument, the dosing device, the sensor
and the stirrer are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead
If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.