5 signatures, Electronic signature, Signatures – Metrohm tiamo 2.4 Manual User Manual
Page 1335
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.4
■■■■■■■■
1319
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)
Comment on modification of determinations
If this check box is activated, then each time that a determination is
modified a modification reason and a modification comment must be
entered that will then be saved in the determination and displayed in the
database in the Information subwindow on the Determination tab.
The reason and comments are also recorded in the Audit Trail.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)
Comment on modification of sample data (live)
If this check box is activated then each time that a determination is
modified a modification reason and a modification comment must be
entered; these are saved in the determination and shown in the database
in the Information subwindow on the Sample tab. The reason and
comments are also recorded in the Audit Trail.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)
6.2.2.5
Signatures
Tab: Configuration
▶ Tools ▶ Security settings ▶ Signatures
The parameters for the electronic signature can be configured on the Sig-
natures tab.
Inactivity delay
If this check box is activated, then the dialog window for signing will be
automatically closed when the entered time limit has expired.
Selection
on | off
Default value
off
Input range
1 - 60 min
Default value
10 min
Remove password after signature
If this check box is activated, then the password must be entered again
after each signature.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)