661
Titrino
Input range
-170.0 - 500.0 °C
Default value
25.0 °C
5.6.3.3.5.5
MET Upol - Stop conditions
Tab: Method
▶ MET Upol ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Entry
25 characters
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.
Stop volume
Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Titrando, 855
Input range
0.00000 - 9999.99 mL
Default value
100,000 mL
Selection
off
Titrino
Input range
0.000 - 9999.99 mL
Default value
100.00 mL
Selection
off
Stop measured value
Stops when the preset value for a measuring point has been exceeded or
not achieved since the start of the titration.
Input range
-200.0 - 200.0 µA
Selection
off
Stop EP
The titration is stopped when the specified number of equivalence points
has been found.
Input range
1 - 9
Default value
9
Selection
off