6 dosing commands, 1 dosing commands - overview, Dosing commands 8 – Metrohm tiamo 2.4 Manual User Manual
Page 1174
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1158
■■■■■■■■
tiamo 2.4
Titrando, 855, 867
Input range
0.1 - 999999.0 s (Increment: 0.1)
Default value
2.0 s
Titrino
Input range
0.08 - 16200 s (Increment: 0.08)
Default value
2.0 s
Temperature
Temperature
The temperature at which calibration is carried out can be entered man-
ually. If a temperature sensor is connected and the Temperature mea-
surement is set on the General/Hardware tab under Sensor to auto-
matic or continuous, then the temperature will be measured continu-
ously. The value is used for temperature correction with the pH measure-
ment.
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
5.6.6
Dosing commands
5.6.6.1
Dosing commands - Overview
Menu item: Method
▶ Insert ▶ New command... ▶ Dosing
Commands for dosing with dosing units or exchange units.
The following dosing commands can be selected:
■
ADD
Dosing of a predefined volume.
■
DOS pH
Controlled dosing with monitoring of measured value pH and tempera-
ture.
■
DOS U
Controlled dosing with monitoring of measured value U and tempera-
ture.
■
LQH
Extensive dosing possibilities with a Dosino.
■
PREP
Preparing an exchange or dosing unit.
■
EMPTY
Empties a dosing unit.
■
RLS DOS
Releases a dosing unit.