5 set ph - titration parameters, Titration parameters – Metrohm tiamo 2.4 Manual User Manual
Page 725
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
709
Titrino
Input range
-20.00 - 20.00
Selection
off
Default value
off
For additional parameters for Endpoint 2, see Endpoint 1.
5.6.3.4.3.5
SET pH - Titration parameters
Tab: Method
▶ SET pH ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Entry
25 characters
Parameters for the run of the titration.
Titration direction
Selection of the titration direction. If two endpoints are set, then the titra-
tion direction is already established and the following selection of titra-
tion direction is ignored.
Selection
+ | - | auto
Default value
auto
+
positive measured value alteration
-
negative measured value alteration
auto
The titration direction is determined automatically from the start mea-
sured value and the set endpoint.
Extraction time
The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). For example, it makes sense to
enter an extraction time for the titration of samples that are not easily
soluble.
Titrando, 719, 720, 736, 751, 758, 785, 794, 798, 799
Input range
0 - 999999 s
Temperature
Manually entered titration temperature. If a temperature sensor is con-
nected and the Temperature measurement is set on the General/
Hardware tab under Sensor to automatic or continuous, then the
temperature will be measured continuously. This value is used for temper-
ature correction in pH measurements.