Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Input range
-15 - 15
Default value
8
Switch on/off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command.
5.6.3.7.3
BRC - Start conditions
Tab: Method
▶ BRC ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Entry
25 characters
Initial measured value
NOTE
Only displayed for Titrando and 855.
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:
Measured value drift
The measured value is only accepted if the drift is less than the value
entered here.
Input range
0.1 - 999.0 mV/min
Selection
off
Default value
off
off
The measured vales will not be applied until after the maximum wait-
ing time has passed.
Min. waiting time
The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting