5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
820
■■■■■■■■
tiamo 2.4
auto
The titration direction is determined automatically from the initial mea-
sured value and the set endpoint.
Extraction time
The titration is not canceled until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, for instance, for the titration of sparingly soluble sam-
ples.
Input range
0 - 999,999 s
Temperature
Manually entered titration temperature. If a temperature sensor is con-
nected and the Temperature measurement is set on the General/
Hardware tab under Sensor to automatic or continuous, then the
temperature will be measured continuously. This value is used for temper-
ature correction in pH measurements.
Titrando, 855
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
Titrino
Input range
-170.0 - 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando
Input range
0.1 - 999,999 s
Default value
2.0 s
Titrino
Input range
0.08 - 16,200 s
Default value
2.00 s
5.6.3.5.3.6
KFT Upol - Stop conditions
Tab: Method
▶ KFT Upol ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Entry
25 characters
Conditions for stopping the titration. If several stop conditions are set,
then the criterion which is fulfilled first will stop the titration.