Metrohm tiamo 2.4 Manual User Manual
Page 534
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
518
■■■■■■■■
tiamo 2.4
Titrando (without 888)
Selection
1 | 2
Default value
1
855, 888
Selection
1
Default value
1
Sensor
Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.
Selection
Sensor name | Metal electrode
Default value
Metal electrode
I(pol)
The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.
Titrando, 855
Input range
–125.0 - 125.0 µA (Increment: 0.5)
Default value
5.0 µA
Titrino
Input range
–127 - 127 µA (Increment: 1)
Default value
5 µA
Electrode check
on | off (Default value: off)
If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.
Temperature measurement
Type of temperature measurement.
Titrando, 855
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.