3 cal meas conc - measuring parameters – Metrohm tiamo 2.4 Manual User Manual
Page 1136
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1120
■■■■■■■■
tiamo 2.4
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
under Temperature will be used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature will be used.
Stirrer
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Input range
-15 - 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this check box is activated, the stirrer will be switched off automatically
when the command has finished.
5.6.5.7.3
CAL MEAS Conc - Measuring parameters
Tab: Method
▶ CAL MEAS Conc ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
The measuring parameters for the calibration are defined on this tab.