5 Method
tiamo 2.4
■■■■■■■■
1107
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Max. number of entries?
712
Input range
0.4 - 999999 s
Default value
2.0 s
856
Input range
0.1 - 999999.0 s
Default value
2.0 s
Temperature
Measurement temperature
Temperature of the solution to be measured. If a temperature sensor is
connected, the temperature entered here is ignored.
Input range
-20.0 - 150.0 °C
Default value
20.0
5.6.5.4.4
CAL Cond - Conductivity standard
Tab: Method
▶ CAL Cond ▶ Properties... ▶ Conductivity standard
Command name
Name of the command.
Entry
25 characters
Definition of the conductivity standard
Selection of the type as to how the data are to be defined for the conduc-
tivity standard.
Selection
Input | Request | Template
Default value
Input
Input
If this option is selected, then the values for the conductivity standard
can be entered directly.
Request
If this option is selected, then first the current temperature of the solu-
tion is measured after the start of the determination. Afterwards, the
dialog window Data Conductivity standard will appear (see Chap-
ter 5.6.5.4.5, page 1108). The method continues running after the
data has been entered.