4 meas u - evaluations, Evaluations – Metrohm tiamo 2.4 Manual User Manual
Page 976
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
960
■■■■■■■■
tiamo 2.4
off
No stop.
Temperature
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.
Titrando, 855, 867
Input range
-20.0 - 150 °C
Default value
25.0 °C
Titrino
Input range
-170.0 - 500 °C
Default value
25.0 °C
5.6.4.4.4
MEAS U - Evaluations
Tab: Method
▶ MEAS U ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Entry
25 characters
The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.
Fixed endpoint evaluation
Fixed endpoint evaluation
on | off (Default value: off)
If this option is activated, then the associated values for the other mea-
sured quantities will be interpolated with a fix measured quantity (Mea-
sured value or Time) for the fixed endpoint from the measuring point
list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).