5 Method
tiamo 2.4
■■■■■■■■
1131
Moves the concentration downward (modifies sequence).
[New]
Add the concentration for the next standard.
[Properties]
Edit the concentration of the selected standard.
[Delete]
Deletes the selected concentration.
Request for standard exchange
on | off (Default value: on)
If this check box is activated, a request for manual solution exchange is
displayed after each time the calibration loop is run through. This parame-
ter must be disabled for automatic solution exchange when calibrating
with a sample changer.
Calibration curve
Calibration function
Selection of the function with which the calibration curve is to be calcu-
lated out of the calibration points.
Selection
linear | quadratic | cubic | automatic
Default value
linear
Confidence interval
Range within which the measured values for the determination of the cali-
bration curve must lie. Values outside this range are called outliers; they
are not integrated in the calculation of the calibration coefficient.
Selection
99% | 98% | 95% | off
Default value
98%
off
The confidence interval will not be checked.
5.6.5.11
CAL MEAS Opt
5.6.5.11.1
CAL MEAS Opt - Overview
Dialog window: Method
▶ CAL MEAS Opt ▶ Properties... ▶ CAL MEAS Opt - 'Com-
mand name'
Command for measuring standard solutions for the calibration of
colorimetric sensors, used together with the CAL LOOP Opt com-
mand.