Metrohm tiamo 2.4 Manual User Manual
Page 530
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
514
■■■■■■■■
tiamo 2.4
Identification
Description
.EP{x}.TEM
Temperature for the endpoint × (1 - 9) in °C
.EP{x}.TIM
Time in s until the endpoint × (1 - 9) is reached
.EP{x}.VOL
Volume for the endpoint × (1 - 9) in mL
.ETE
End temperature (temperature after the command
has been processed) in °C
.EVT
End volume (total dosed volume at the end of the
command) in mL
.FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) =
Command has never ended
.FP{x}.ERC
ERC for the fixed endpoint × (1 - 9)
.FP{x}.MEA
Measured value for the fixed endpoint × (1 - 9) in mV
.FP{x}.TEM
Temperature for the fixed endpoint × (1 - 9) in °C
.FP{x}.TIM
Time in s until the fixed endpoint × (1 - 9) is reached
.FP{x}.VOL
Volume for the fixed endpoint × (1 - 9) in mL
.GP.VOL
Volume for the Gran endpoint in mL
.GP.MEA
Measured value for the Gran endpoint in the unit of
the measured value
.GP.TEM
Temperature for the Gran endpoint in °C
.GP.TIM
Time in s until the Gran endpoint is reached
.HP{x}.MEA
Measured value for the HNP × (1 - 9) in mV (HNP =
half neutralization potential)
.HP{x}.TEM
Temperature for the HNP × (1 - 9) in °C
.HP{x}.TIM
Time in s until the HNP × (1 - 9) is reached
.HP{x}.VOL
Volume for the HNP × (1 - 9) in mL
.IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
.ITE
Initial temperature (temperature before start condi-
tions are processed) in °C
.LP.CAx
Calculated value × (1 - 3) for the last measuring point
in the measuring point list