Metrohm tiamo 2.4 Manual User Manual
Page 1333
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.4
■■■■■■■■
1317
Password policies
NOTE
If the Password monitoring by Windows option is enabled, then
the parameters for password protection are deactivated. Compliance
with the directives pursuant to 21 CFR Part 11 must then be ensured by
the administrator in Windows.
Enforce unique password
Activating this check box ensures that a user can use a given password
only once.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)
Enforce use of special characters
Activating this check box ensures that the password must contain at
least one special character (@, #, ~, etc.).
Selection
on | off
Default value
off
Minimum password length
Activating this check box ensures that the password must contain at
least the indicated number of characters.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)
Input range
1 - 10 characters
Default value
6 characters
Passwords expire every
Activating this check box ensures that the user must enter a new pass-
word before the validity period expires. When a user logs in whose pass-
word will expire within the next 10 days, a corresponding message will
appear. If the validity period has expired, the user can only log in if he or
she changes the password.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)