5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
880
■■■■■■■■
tiamo 2.4
since
Specifies when the stop time begins to count (will not be displayed if the
Stop time is disabled):
Selection
Start | Control point first reached | Last dosing
Default value
Start
Start
The stop time starts after the start volume has been dosed and after
the following pause time.
Control point first reached
The stop time starts as soon as the control point has been reached for
the first time.
Last dosing
The stop time starts after the last increment has been dosed, i.e., the
stop time is reset to zero again after each dosing step.
Stop rate
Stop when the current dosing rate is lower than the defined stop rate.
This criterion is not checked until 10 s after the start. The maximum stop
rate depends on the cylinder volume of the exchange unit or dosing unit
used.
Titrando, 855
Input range
0.01 - 166.00 mL/min
Selection
off
Default value
off
Titrino
Input range
0.01 - 150.00 mL/min
Selection
off
Default value
off
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.
Titrando, 855
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Input range
0.01 - 150.00 mL/min