Metrohm tiamo 2.4 Manual User Manual
Page 817
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
801
Titrando
Input range
1 - 999 µg/min
Default value
20 µg/min
Drift correction
If this parameter is set to automatic or manual the drift correction ( =
Drift value × Time for drift correction) caused by empty consumption
during titration is deducted from the total consumption. The Time for
drift correction is the period between end of conditioning and end of
titration.
Selection
automatic | manual | off
Default value
off
automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Input range
0.0 - 99.9 µL/min
Default value
0.0 µL/min
Stop volume
Stop of the determination when the specified volume has been dosed
since the start of the conditioning. The stop volume should be adjusted to
the size of the titration cell in order to prevent any overflow.
Titrando
Input range
0.000000 - 9,999.99 mL
Default value
20.000 mL
Selection
off
Default value
off
Stop time
Stop of the determination when the specified time has elapsed since the
start of the conditioning.
Titrando
Input range
1 - 999,999 s