5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1002
■■■■■■■■
tiamo 2.4
Lower limit
Value for the lower limit of the window.
Time for the lower limit of the window.
Window = Measured value
Input range
-20.0 - 150.0 °C
Default value
-20.0 °C
Window = Time
Input range
0 - 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.
Value for the upper limit of the window.
Window = Measured value
Input range
-20.0 - 150.0 °C
Default value
150.0 °C
Window = Time
Input range
1 - 999999 s
Default value
999999 s
5.6.4.7.5
MEAS T - Additional measured values
Tab: Method
▶ MEAS T ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Entry
25 characters
A maximum of 6 additional measured values can be defined here which
can then be saved together with the measured values present in the
default settings in additional measured value columns Calculated # or
External # in the measuring point list.
Additional calculated measured values
Additional calculated measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be
presented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.