Metrohm tiamo 2.4 Manual User Manual
Page 1251
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1235
If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
774 Oven Sample Processor | 089 Photome-
ter | 874 Oven Sample Processor
Default value
874 Oven Sample Processor
Heater (Sample changer)
Target temperature
Temperature to which the oven has to be heated. This value is live edita-
ble only for the 874.
Input range
50 - 250 °C
Selection
Init.temp. | off
Default value
Init.temp.
Init.temp.
With Init.temp. the oven heating is switched on at the same time as
the device and the oven is heated up to this temperature. The initial
temperature must be entered by keypad on the 774. When connecting
to tiamo, the value is read out and displayed in the device configura-
tion. The value cannot be altered in tiamo. The initial temperature is
entered and altered in tiamo with the 874.
off
The heater is switched off during the determination.
NOTE
It is advisable to define an initial temperature if you work always at the
same temperature.
Heating time
The target temperature has to be reached within this time. The heating
rate can be modified with the heating time in order to program tempera-
ture ramps or entire temperature profiles (with several HEATER com-
mands). Off causes heating at the maximum possible heating rate (see
device manual).
This value is live editable for the 874
Input range
1 - 999 min
Selection
off
Default value
off