4 stdadd auto - measuring parameters – Metrohm tiamo 2.4 Manual User Manual
Page 1109
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1093
Selection
mol/L | mmol/L | µmol/L | ppm | g/L | mg/L |
µg/L | mg/mL | µg/mL | % | mEq/L
Default value
ppm
Control parameters
Dosing rate
Speed at which the volume increments are added.
Selection
slow | medium | fast
Default value
fast
Delta U
Potential difference that is to be achieved by a standard addition.
Input range
1 - 999 mV
Default value
10 mV
5.6.4.19.4
STDADD auto - Measuring parameters
Tab: Method
▶ STDADD ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement with drift control
The measurement is stopped as soon as the defined Signal drift or the
Stop measured value is reached or a defined Waiting time has
elapsed.
Measured value drift
The measured value is only accepted if the signal drift defined here has
been fallen short of.
Input range
0.1 - 999.0 mV/min
Default value
10.0 mV/min
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the waiting time is elapsing.