9 How to proceed?
tiamo 2.4
■■■■■■■■
1631
The database restoring is started.
NOTE
Existing databases cannot be overwritten, i.e. they must first be
deleted so that the database can be recovered under its old name.
9.2.3
Backing up configuration data
How to proceed?
General
The configuration data is saved in tiamo in the Configuration data-
base. Configuration data includes all settings that apply to all methods,
i.e. settings for devices, titrants/solutions, sensors, common variables and
rack data as well as Methods, Security settings (see Chapter 6.2.2.1,
page 1312), User administration (see Chapter 6.2.1.1, page 1303),
Program administration (see Chapter 6.2.3.1, page 1322), templates
and Audit Trail (see Chapter 6.4, page 1352).
In local server systems (tiamo light, tiamo full), the configuration data-
base is located in the program directory of the computer on which the
program has been installed. In client/server systems (tiamo multi), the
configuration database is stored centrally on the server and contains all
the configuration data of all computers (clients) that are connected to this
server.
NOTE
It is strongly recommended that the configuration database be backed
up periodically.
Backing up configuration data manually
1 Select the Configuration program part.
2 Click on the File
▶ Backup ▶ Manually menu item.
The Backup configuration data manually dialog window opens.
3 Select the directory for the backup in the Backup directory field.
4 Select or enter a new name for the backup file in the Backup name
list box. If an existing backup file is selected, it will be overwritten.