Metrohm tiamo 2.4 Manual User Manual
Page 1068
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1052
■■■■■■■■
tiamo 2.4
Max. waiting time
If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered, then a waiting time that is suitable for
the drift will be calculated automatically according to the following for-
mula:
Input range
0 - 999,999 s
Default value
154 s
Measuring interval
Time interval for entering a measured value in the measuring point list.
Avantes Spectrometer
Input range
0.1 - 999,999.0 s (Increment: 0.1)
Default value
2.0 s
089 Photometer
Input range
0.25 - 999,999.00 s (Increment: 0.25)
Default value
2.0 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Input range
0 - 4,000 mAU
Selection
off
Default value
off
Measurement without drift control
If the Measurement without drift control option has been selected,
then the following parameters will be displayed:
Measuring time
Duration of the absorbance measurement.
Input range
0 - 999,999 s
Default value
120 s
Measuring interval
Time interval for entering a measured value in the measuring point list.
Avantes Spectrometer
Input range
0.1 - 999,999.0 s (Increment: 0.1)
Default value
2.0 s