8 brc - additional evaluations, Additional evaluations – Metrohm tiamo 2.4 Manual User Manual
Page 873
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
857
NOTE
Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titra-
tion will be ignored.
If this command is already started (e.g. via Automatic start after
sample addition), the SEND command will not have any effect.
Show measured value during conditioning
on | off (Default value: off)
If this option is enabled, then the measured value will be shown in the
subwindow Live display of the program part Workplace, in addition to
the drift and the time.
5.6.3.7.8
BRC - Additional evaluations
Tab: Method
▶ BRC ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Entry
25 characters
The following additional method for evaluation of the titration curves can
be activated and defined on this tab:
Fixed endpoint evaluation
Fixed endpoint evaluation
on | off (Default value: off)
If this option is activated, then the associated values for the other mea-
sured quantities will be interpolated with a fixed measured quantity (mea-
sured value, time or bromine) for the fixed endpoint from the measur-
ing point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.7.10, page 859).