4 meas tc cond - additional measured values – Metrohm tiamo 2.4 Manual User Manual
Page 1052
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1036
■■■■■■■■
tiamo 2.4
Start temperature
Temperature at the start of the measurement.
Input range
-20.0 - 150.0 °C
Default value
20.0 °C
Stop temperature
Temperature at the time of the end of the measurement.
856
Input range
-20.0 - 150.0 °C
Default value
40.0 °C
5.6.4.11.4
MEAS TC Cond - Additional measured values
Tab: Method
▶ MEAS U ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Entry
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together with the measured values present in
the default settings in additional measured value columns Calculated #
or External # in the measuring point list.
Additional calculated measured values
Additional calculated measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be
presented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.
[New]
Open the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be entered (see Chapter
5.6.3.11.1, page 933).
[Properties]
Open the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be edited (see Chapter
5.6.3.11.1, page 933).
[Delete]
Delete the calculated measured value selected in the table.