5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
620
■■■■■■■■
tiamo 2.4
5.6.3.3.3.11
MET U - Fixed endpoint evaluation
Dialog window: Method
▶ MET U ▶ Properties... ▶ Additional evaluations ▶ Fixed
endpoint evaluation
▶ [New]/[Properties] ▶ Fixed endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Input range
-2000.0 - 2000.0 mV
Time
Input range
0.0 - 999999.9 s
Volume
Input range
0.00000 - 9999.99 mL
5.6.3.3.4
MET Ipol
5.6.3.3.4.1
MET Ipol - Overview
Dialog window: Method
▶ MET Ipol ▶ Properties... ▶ MET Ipol - 'Command name'
Command for Monotonic equivalence point titrations with voltamet-
ric measurement (selectable polarization current).
Instruments
This command can be executed with the following devices.
Titrando: 808, 809, 835, 836, 852, 857, 888, 904, 905, 906, 907
Titrino: 702, 716, 721, 736, 751, 785, 794, 798, 799
Robotic Titrosampler: 855
Appearance
The command has the following appearance: