5 Method
tiamo 2.4
■■■■■■■■
509
5.6.3.2.3.9
DET U - Measured value window
Dialog window: Method
▶ DET U ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with measured value window (U) ▶ [New/[Properties] ▶ Mea-
sured value window #
Measured value windows are regions (windows) on the measured value
axis for which different parameters for potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-
tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.
Lower limit
Lower limit for the measured value window.
Titrando, 855
Input range
-2000.0 - 2000.0 mV
Default value
-2000.0 mV
Titrino
Input range
-2000 - 2000 mV
Default value
-2000 mV
Upper limit
Upper limit for the measured value window.
Titrando, 855
Input range
-2000.0 - 2000.0 mV
Default value
2000.0 mV
Titrino
Input range
-2000 - 2000 mV
Default value
2000 mV
EP criterion
Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.
Titrando, 855
Input range
0 - 200
Default value
5
EP recognition
Filter for the recognition of equivalence points: