5 Method
tiamo 2.4
■■■■■■■■
1181
5.6.6.3.4.5
DOS pH - Monitoring
Tab: Method
▶ DOS pH ▶ Properties... ▶ Monitoring
Command name
Name of the command.
Entry
25 characters
The following monitoring processes can be switched on and defined on
this tab:
Monitoring Measured value
Monitoring Measured value
on | off (Default value: off)
If this option is switched on then the measured value will be monitored
and any exceeded limits will be entered in the measuring point list.
Lower limit pH
Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is trig-
gered.
Titrando, 855
Input range
-20.000 - 20.000
Default value
-20.000
Titrino
Input range
-20.00 - 20.00
Default value
-20.00
Lower hysteresis pH
Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value
lower limit ok is triggered.
Titrando, 855
Input range
0 - 20.000
Default value
20
Upper limit pH
Upper limit of the measured value. If the measured value exceeds this
limit, then the event Measured value upper limit exceeded is trig-
gered.