2 det u - general/hardware – Metrohm tiamo 2.4 Manual User Manual
Page 505
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
489
Identification
Description
.SLO
Electrode slope of the sensor used for the command
(in mV for ISE sensors)
.SME
Start measured value (measured value after process-
ing the start conditions) in the unit of the measured
value
.STE
Start temperature (temperature after processing the
start conditions) in °C
.STY
Type of stop with which the command was stopped:
1 = normal; 0 = manual or after error
.SVA
Start volume absolute (volume that was added
according to the start condition "start volume") in mL
.SVM
Start volume measured value (volume that was added
according to the start condition "start measured
value") in mL
.SVS
Start volume slope (volume that was added according
to the start condition "start slope") in mL
.SVT
Total start volume (volume that was added according
to all three start conditions) in mL
.TITER
Titer value of the solution used for the command
5.6.3.2.3.2
DET U - General/Hardware
Tab: Method
▶ DET U ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the instrument, the dosing device, the sensor
and the stirrer are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined