5 Method
tiamo 2.4
■■■■■■■■
617
Additional external measured values
Additional external measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can
be taken over and saved from existing measuring commands (i.e. MEAS)
which can also be displayed in curves under the designation Extern 1…3
and used in formulas as variables 'Command name.EX1…3'-
[New]
Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 934).
[Properties]
Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 934).
[Delete]
Delete the external measured value selected in the table.
5.6.3.3.3.9
MET U - Measured value window
Dialog window: Method
▶ MET U ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with measured value window ▶ [New]/[Properties] ▶ Mea-
sured value window#
Measured value windows are regions on the measured value axis for
which different parameters for potentiometric evaluation can be defined.
Only those endpoints which lie within these windows and additionally
meet the parameters defined for each window will be recognized. Only
one endpoint will be recognized per window.
Lower limit
Lower limit for the measured value window.
Titrando, 855
Input range
-2000.0 - 2000.0 mV
Default value
-2000.0 mV
Titrino
Input range
-2000 - 2000 mV
Default value
-2000 mV
Upper limit
Upper limit for the measured value window.