Metrohm tiamo 2.4 Manual User Manual
Page 940
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
924
■■■■■■■■
tiamo 2.4
Selection
off
Default value
off
Stop slope
Stop when the criterion defined for the first derivative of the measured
value (slope) is fulfilled.
First field: Selection of the reference condition.
Selection
< | >
Default value
<
Second field: Definition of reference value for the derivative of the mea-
sured value.
Input range
-999999.999 - 999999.999 °C/mL (Incre-
ment: 0.001)
Default value
0.100 °C/mL
Selection
off
Stop slope active after
Required minimum volume for the stop. The field Stop slope active
after is active only if the comparison value is not set to off for Stop
slope.
Input range
0.00 - 50.00 mL
Selection
off
Default value
off
Additional volume after stop
After stopping the titration, this additional volume is dosed according to
the stop conditions defined. The curve shape after reaching the endpoint
can also be seen this way.
Input range
0.01 - 50.00 mL
Default value
0.5 mL
Selection
off
Filling rate
Rate (volume/time unit) at which the buret is to be filled after the titration.
The maximum filling rate depends on the cylinder volume of the dosing
unit used. If the selected filling rate is higher than the maximum possible
dosing rate, it will be automatically corrected to this value.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum