9 set upol - additional measured values, Additional measured values – Metrohm tiamo 2.4 Manual User Manual
Page 797
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
781
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the mea-
suring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.5 - 10.0 µA/mL
Default value
5.0 µA/mL
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the mea-
suring point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.5 - 10.0 µA/mL
Default value
5.0 µA/mL
5.6.3.4.6.9
SET Upol - Additional measured values
Tab: Method
▶ SET Upol ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Entry
25 characters
A maximum of 6 additional measured values can be defined on this tab.
These values can then be saved together in additional measured value col-
umns with the measured values present in the default settings.