10 brc - fixed endpoint evaluation – Metrohm tiamo 2.4 Manual User Manual
Page 875
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
859
which can also be displayed in curves under the designation Extern 1…3
and used in formulas as variables 'Command name.EX1…3'-
[New]
Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 934).
[Properties]
Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 934).
[Delete]
Delete the external measured value selected in the table.
5.6.3.7.10
BRC - Fixed endpoint evaluation
Dialog window: Method
▶ BRC ▶ Properties... ▶ Additional evaluations ▶ Fixed
endpoint evaluation
▶ [New]/[Property] ▶ Fixed endpoint evaluation #
Measured quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Bromine
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Input range
-2000.0 - 2000.0 mV
Time
Input range
0.0 - 999999.9 s
Bromine
Input range
0.0 - 999999.9 µg