Metrohm tiamo 2.4 Manual User Manual
Page 872
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
856
■■■■■■■■
tiamo 2.4
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Input range
1 - 999999 s
Selection
off
Default value
off
Stabilizing time
Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.
Input range
0 - 999 s
Default value
0 s
Automatic start after sample addition time
on | off (Default value: on)
If this option is activated, then the titration will be started automatically
after the sample addition time.
Sample addition time
Waiting time between stop of conditioning and start of titration during
which the sample can be added.
Input range
0 - 99 s
Default value
0 s
Manual start after sample addition with [Continue]
on | off (Default value: off)
If this option is activated, then the titration will be started manually after
the sample addition by pressing the [Continue] button in the message
window.
Only start titration by a start command from a SEND command
on | off (Default value: off)
If this option is activated, then the titration will not be started until the
Event message Start titration is sent to the command with a SEND
command.