3 elt meas - measuring parameters – Metrohm tiamo 2.4 Manual User Manual
Page 1144
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1128
■■■■■■■■
tiamo 2.4
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will be used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 867
Selection
1 | 2 | 3 | 4
Default value
1
855
Selection
1 | 2 | 3
Default value
1
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Input range
-15 - 15 (without 0)
Default value
8
5.6.5.9.3
ELT MEAS - Measuring parameters
Tab: Method
▶ ELT MEAS ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
Temperature
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.
Input range
-20.0 - 150 °C
Default value
25.0 °C