3 cal meas ph - measuring parameters – Metrohm tiamo 2.4 Manual User Manual
Page 1172
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1156
■■■■■■■■
tiamo 2.4
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
under Temperature will be used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature will be used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855, 867
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855, 867
Input range
-15 - 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos, 855 and 867.
5.6.5.13.3
CAL MEAS pH - Measuring parameters
Tab: Method
▶ CAL MEAS pH ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
The measuring parameters for the calibration are defined on this tab.