Metrohm tiamo 2.4 Manual User Manual
Page 1566
7.12 712 Conductometer
■■■■■■■■■■■■■■■■■■■■■■
1550
■■■■■■■■
tiamo 2.4
GLP test interval
Time interval to next GLP test. If a value is entered here then the date in
the field Next GLP test will be adjusted automatically.
Can only be edited if monitoring is switched on.
Input range
1 - 999 days
Default value
999 days
Next GLP test
Date on which the next GLP test is to be carried out. The date can be
selected by pressing
in the dialog window Select date (see Chapter
2.5.1, page 90). After the date has been entered the field GLP test
interval will be adjusted automatically.
Can only be edited if monitoring is switched on.
Default value
GLP test date + 999 days
Message
The message options can only be edited if monitoring is switched on.
Message by e-mail
on | off (Default value: off)
If this option is activated, the message is sent to the e-mail address
defined under [E-mail...]l. The message is sent in text format.
[E-mail...]
[E-mail...] opens the Send e-mail window (see Chapter 2.6.1, page 93).
Acoustic signal
on | off (Default value: off)
If this option is activated, an acoustic signal will be emitted in addition to
the message defined above.
Action
The actions can only be edited if monitoring is switched on.
If during monitoring it is found that the validity period has expired then
one of the following actions will be triggered automatically at start test:
Selection
Record message | Display message | Cancel
determination
Default value
Display message
Record message
The message that the validity period has expired will be automatically
saved with the determination.